<?xml version="1.0" encoding="UTF-8"?>
|
|
<?oxygen RNGSchema="http://www.oasis-open.org/docbook/xml/5.0/rng/docbook.rng" type="xml"?>
|
|
<chapter xmlns="http://docbook.org/ns/docbook" xmlns:xlink="http://www.w3.org/1999/xlink"
|
|
version="5.0">
|
|
<info><title>Ordela Detector Voltage Control</title><author>
|
|
<personname>Ferdi Franceschini</personname>
|
|
</author>
|
|
<date>2008-08-29 16:47</date>
|
|
</info>
|
|
<sect1>
|
|
<title>Commands</title>
|
|
<para>The High Voltage controller for the Ordela detector has been implemented as a standard
|
|
SICS environment controller object with a driveable interface. It has been configured
|
|
differently to other SICS objects in several ways. Firstly, you use
|
|
<command>up</command> and <command>down</command> commands to drive the voltage to its
|
|
<command>upper</command> and <command>lower</command> limits. This is a blocking
|
|
task i.e. no other task can started until this is complete. Secondly, the instrument has
|
|
been configured with the SICS anticollider to prevent you from moving the detector when
|
|
the voltage is above a certain threshold, which will lead to damage of the detector.
|
|
This is important for Quokka as the detector is moved frequently. </para>
|
|
<variablelist>
|
|
<varlistentry>
|
|
<term><command>dhv1 up</command></term>
|
|
<listitem>
|
|
<para>Raise the voltage</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 down</command></term>
|
|
<listitem>
|
|
<para>Lower the voltage</para>
|
|
<note>
|
|
<para>NOTE This command blocks until the power supply reaches the "upper" or
|
|
"lower" running voltages, see below.</para>
|
|
</note>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>INT1712 1</command></term>
|
|
<listitem>
|
|
<para>If this commands hang SICS you can interrupt it with by entering this at
|
|
the SICS command line, or by pressing the interrupt button at the bottom of
|
|
GumTree</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 reset</command></term>
|
|
<listitem>
|
|
<para>Reset the controller </para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 list</command></term>
|
|
<listitem>
|
|
<para>Displays the values of the various parameters</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 send </command><replaceable>command</replaceable></term>
|
|
<listitem>
|
|
<para>Sends a <replaceable>command</replaceable> to the unit and displays the
|
|
response</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 off </command></term>
|
|
<listitem>
|
|
<para>Drives the output voltage to zero</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
</variablelist>
|
|
</sect1>
|
|
<sect1>
|
|
<title>Parameters</title>
|
|
<variablelist>
|
|
<varlistentry>
|
|
<term><command>dhv1 upper </command><replaceable>voltage</replaceable></term>
|
|
<listitem>
|
|
<para>Sets the running voltage for the <command>up</command> command. This would
|
|
normally be the operating voltage for the equipment to which the power
|
|
supply is connected.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 lower </command><replaceable>voltage</replaceable></term>
|
|
<listitem>
|
|
<para>Sets the standby voltage for the <command>down</command> command. This
|
|
would normally be the standby voltage for the equipment to which the power
|
|
supply is connected.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 max </command><replaceable>voltage</replaceable></term>
|
|
<listitem>
|
|
<para>Sets the hardware maximum for the power supply. For the Ordela power
|
|
supplies, it is important that this is the correct full-scale value of the
|
|
power supply itself. This is used to convert between the voltage step and
|
|
voltages and to calculate the step period from the voltage slew rate.</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 rate </command><replaceable>voltage</replaceable></term>
|
|
<listitem>
|
|
<para>The volts per second at which the power supply slews between voltages. For
|
|
the Ordela power supplies, this is used to calculate the time between
|
|
voltage steps based on the <option>max</option> parameter</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 debug </command><replaceable>val</replaceable></term>
|
|
<listitem>
|
|
<para>Allowed <replaceable>val</replaceable></para>
|
|
<para><option>0</option> No debug information in log</para>
|
|
<para><option>1</option> Debug information in log</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 lock </command></term>
|
|
<listitem>
|
|
<para>This locks the device from being set by users. Users can use <command>up
|
|
down</command> and <command>off</command> commands to set
|
|
voltages</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
<varlistentry>
|
|
<term><command>dhv1 unlock </command></term>
|
|
<listitem>
|
|
<para>Managers may unlock the device</para>
|
|
</listitem>
|
|
</varlistentry>
|
|
</variablelist>
|
|
</sect1>
|
|
</chapter>
|