24 lines
750 B
Tcl
24 lines
750 B
Tcl
# Detector voltage controller
|
|
|
|
set sim_mode [SplitReply [detector_simulation]]
|
|
|
|
if {$::sim_mode == "true"} {
|
|
EvFactory new dhv1 sim
|
|
} else {
|
|
# clientput "Detector Voltage control not yet available"
|
|
# MJL 6/13 Revised voltage settings for serviced original Platypus detector (rectangular DENEX 250TN)
|
|
# Old voltages +3350V and -1500V, after servicing gas pressure changed and now +2800V and -1000V.
|
|
# NOTE: If square DENEX 300TN is fitted, voltages need to be revised again! (+2850V and -1100V currently for this detector)
|
|
makeasyncqueue acq NHQ200 ca2-platypus 4001
|
|
evfactory new dhv1 nhq200 acq 1
|
|
dhv1 tolerance 5
|
|
dhv1 rate 25
|
|
dhv1 upper 2800
|
|
#dhv1 lock
|
|
evfactory new dhv2 nhq200 acq 2
|
|
dhv2 tolerance 5
|
|
dhv2 rate 25
|
|
dhv2 upper 1000
|
|
#dhv2 lock
|
|
}
|