Low Frequency Amplifier. LF AG1010Jing Chen2008-08-29 16:47ConfigurationThe device LF AG1010 is connected to a Moxa 5430 server. SICS client communicates with
the Moxa 54530 server who sends and receives data packets to/from the LF AG1010 through
a RS232 serial interface. Settings of the Moxa 5430 areBits Per Second – 19200Data Bits – 8Parity – NONE Stop Bit – 1Flow Controls – NONEPort Number – 4001.CommandsCommands for getting and setting values on the LF AG1010. Sending a command without a
value will return the current value. Note that all power values are in deciwatts (dW). 1dW = 0.1Watt. lf_limits powerSet forward power limit to power dW.lf_limits FPL 50 sets the forward power limit to 50dWlf_limits FPL returns the forward power limitUnits: dWRange: 0 to 6000lf_limits powerSet reverse power limit to power dWlf_limits RPL 50 sets the reverse power limit to 50dWlf_limits RPL returns the reverse power limitUnits: dWRange: 0 to 1600lf_pagc powerSet power level for AGC mode to power dWlf_pagc 50 sets the AGC mode power level to 50dWlf_pagc returns the AGC mode power levellf_pmgc powerSet power level for MGC mode to power %MGC scale in % is for reference only. 0% - 0W, 100% full output. Not
linear! This mode is for continues and pulsed operation.lf_pmgc 50 sets the MGC mode power level to 50%lf_pmgc returns the MGC mode power levelUnits: %Range: 0 to 100lf_freq freqSet frequency to freq HzRange: 20000 to 2000000 (20kHz to 2MHz)lf_burst modemode time
time top
topSet burst mode to modeBURST is possible in MGC Mode only!Allowed mode Burst mode OFF Internal burst mode ON Change burst parameters without changing burst ON/OFF Enable external burst modeSet repetition period of burst cycle to timeUnits: msRange: 1 to 50Set time of power in burst cycle to topUnits: μsRange: 1 to 500lf_burst mode 1 rtime 20 top 5Set burst mode to Internal ModeSet repetition period of burst cycle to 20 msSet time of power in burst cycle to 5 μslf_sweep mode mode start start
step step np
npSets the sweep parameters start Hz, with step
size of step Hz and with number of points
npAllowed mode Sweep mode OFF Sweep mode ON Change sweep parameters without changing sweep mode
ON/OFFAllowed step increments of 1000Hzlf_sweep mode 1 start 20000 step 1000 np 200Set sweep mode on, start frequency = 20kHz, step frequency = 1kHz, number
of points = 200lf_sweep_runstart step npThis command triggers a frequency sweep starting at
start Hz, with step of
step Hz and with number of steps =
np in full sweep cycle under optional
If the mode is 0 (OFF) and mode is not provide, the command will set the
mode the to 2 before the sweeping and reset to 0 (OFF)
after the sweeping.Allowed Sweep mode OFF Sweep mode ON Change sweep parameters without changing sweep mode
ON/OFFlf_meas Returns forward power, reverse power and temperature?start Hz ParametersN/A There are no parameters to set on this device.DescriptionThe AG 1010 produces up to 1,000 Watts of power over a frequency range from lower than
20 kHz to higher than 1.2 MHz. It operates over the entire frequency range without band
switching or other adjustments. Extended range to over 2 MHz is possible with reduced
output power. Gain is rated at 60 dB with a typical gain flatness of ±1.5 dB. The Front Panel offers a LCD display of Forward, Reflected and Load Power readings,
RF Status, MGC/AGC setups and operating frequency in Generator Mode. Power meters are calibrated into a 50 Ohm Load and are accurate when unit operates
into matched load. Outside of matched condition, the model AG 1010’s power measurement
system provides an accurate reading of VSWR. When used as amplifier, the AG 1010 is compatible with most signal and function
generators, computer synthesizer cards and accurately reproduces all waveforms within
its output and bandwidth limits.The Forced-air cooling system and the internal power supply are designed to permit
operation over a wide range of temperature and global AC line conditions. The AG 1010 is
built to withstand a +5 dBm (1.2Vp-p) Input signal. The unit amplifies the inputs of AM,
FM, SSB, pulse and other complex modulations. OUTPUT PROTECTION AG 1010 is protected by its internal control system for 1,000 Watts
of total Forward Power and 160 Watts of Reflected Power. This will protect the amplifier
output stage from accidental overdrive at the input and an extreme mismatch at the
Output.Known IssuesNo know issues.