#AUTO NO TE: $Revision: 1.4 $ NO TE: $Date: 2006-11-02 21:23:39 $ NO TE: Author: Dan Bartlett NO TE: Limit switch HOME routine added by Ferdi Franceschini NO TE: Last revision by: $Author: ffr $ NO TE: NO TE: A-MONOCHROMATOR UPPER TILT (mphi) - TILT 1 NO TE: B-MONOCHROMATOR LOWER TILT (mchi) - TILT 2 NO TE: C-MONOCHROMATOR UPPER TRANSLATION (mz:nexus, mx:KDL) - LINEAR 1 NO TE: D-MONOCHROMATOR LOWER TRANSLATION (mx:nexus, my:KDL) - LINEAR 2 NO TE: E-MONOCHROMATOR ROTATION (mom) - ROTATE NO TE: F-FLIGHT TUBE ROTATE NO TE: G-PRIMARY COLLIMATOR 1 NO TE: H-PRIMARY COLLIMATOR 2 MT-2,-2,-2,-2,-2,-2,-2,-2;'CONFIGURES ALL AXES AS STEPPER MO;' FIRST TIME MOTOR OFF CN 1;' CONFIGURES LIMIT SWITCHES TO ACCEPT NC CONTACTS II 8,,,0;' CONFIGURES THE 8TH INPUT TO ACTIVATE THE #ININT ROUTINE OP 65280;' SETS OUTPUTS 9-16 ON = RELAYS OFF IA 137,157,201,3;' CONFIGURES IP ADDRESS AC 25000,25000,25000,25000,25000,10000,25000,25000;' SET ACELERATION DC 25000,25000,25000,25000,25000,10000,25000,25000;' SET DECELERATION SP 25000,25000,25000,25000,50000,100000,25000,25000;' SET SPEED SIA=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SIB=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SIC=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SID=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SIE=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SIF=1,25,25,0<-1>2;'HEIDENHAIN RCN226 SIG=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 SIH=1,25,13,0<-1>2;'HEIDENHAIN ROQ425 DONE=0;' FIRST TIME THROUGH AIRPAD CONTROL NOT DONE JS #SOLCTRL EN #SOLCTRL NO TE: CONTROLS THE AIRPAD SOLENOID VALVES IF (_MOF=0);'FLIGHT TUBE MOTOR IS ON SB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID OFF SB 12;'DRIVE WHEEL UP-SOLENOID OFF CB 9;'SAMPLE STAGE AIR PAD ON-SOLENOID ON CB 11;'DRIVE WHEEL DOWN-SOLENOID ON DONE=0 ENDIF IF (_MOF=1)&(DONE=0);'FLIGHT TUBE MOTOR IS OFF SB 9;'SAMPLE STAGE AIR PAD ON-SOLENOID OFF SB 11;'DRIVE WHEEL DOWN-SOLENOID OFF CB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID ON CB 12;'DRIVE WHEEL UP-SOLENOID ON WT500;'WAIT 1/2 SEC SB 10;'SAMPLE STAGE AIR PAD OFF-SOLENOID OFF SB 12;'DRIVE WHEEL UP-SOLENOID OFF DONE=1 ENDIF JP #SOLCTRL EN #TCPERR NO TE: EMPTY ROUTINE FOR A TCP ERROR RE #ININT NO TE: INTERUPT ROUTINE TO STOP ALL AXES NO TE: WHEN SAFETY & INTERLOCK SYSTEM REQUESTS IT STA,B,C,D,E,F,G,H AM MO RI #LIMSWI NO TE: LIMIT SWITCH ROUTINE WHICH SHORTENS DECEL TIME WHEN A LIMIT IS HIT OLDDCA=_DCA;' REMEMBERS CURRENT DECEL OLDDCB=_DCB OLDDCC=_DCC OLDDCD=_DCD OLDDCE=_DCE OLDDCF=_DCF OLDDCG=_DCG OLDDCH=_DCH IF (_LFA = 0) | (_LRA = 0) DCA=100000;' DECEL FOR LIMIT HIT OCCURENCE STA AMA MOA DCA=OLDDCA;' RESET DECEL TO OLD VALUE ENDIF IF (_LFB = 0) | (_LRB = 0) DCB=100000;' DECEL FOR LIMIT HIT OCCURENCE STB AMB MOB DCB=OLDDCB;' RESET DECEL TO OLD VALUE ENDIF IF (_LFC = 0) | (_LRC = 0) DCC=100000;' DECEL FOR LIMIT HIT OCCURENCE STC AMC MOC DCC=OLDDCC;' RESET DECEL TO OLD VALUE ENDIF IF (_LFD = 0) | (_LRD = 0) DCD=100000;' DECEL FOR LIMIT HIT OCCURENCE STD AMD MOD DCD=OLDDCD;' RESET DECEL TO OLD VALUE ENDIF IF (_LFE = 0) | (_LRE = 0) DCE=100000;' DECEL FOR LIMIT HIT OCCURENCE STE AME MOE DCE=OLDDCE;' RESET DECEL TO OLD VALUE ENDIF IF (_LFF = 0) | (_LRF = 0) DCF=50000;' DECEL FOR LIMIT HIT OCCURENCE STF AMF MOF DCF=OLDDCF;' RESET DECEL TO OLD VALUE ENDIF IF (_LFG = 0) | (_LRG = 0) DCG=100000;' DECEL FOR LIMIT HIT OCCURENCE STG AMG MOG DCG=OLDDCG;' RESET DECEL TO OLD VALUE ENDIF IF (_LFH = 0) | (_LRH = 0) DCH=100000;' DECEL FOR LIMIT HIT OCCURENCE STH AMH MOH DCH=OLDDCH;' RESET DECEL TO OLD VALUE ENDIF RE