- Enhanced and debugged histogram memory for AMOR
* added PROJECT both in HM and driver code * added single detector support. - Removed several bugs in the AMOR data bit. - Updated documentation
This commit is contained in:
@ -1,59 +1,59 @@
|
||||
<HTML>
|
||||
<HEAD>
|
||||
<TITLE>Moving SICS</TITLE>
|
||||
<TITLE>Replacing Hardware</TITLE>
|
||||
</HEAD>
|
||||
<BODY>
|
||||
<H1>Moving SICS</H1>
|
||||
<H2> Moving the SICS Server to a new Computer</h2>
|
||||
<P>
|
||||
Moving the SICS server from one computer to another involves quite a
|
||||
few steps:
|
||||
<OL>
|
||||
<LI>Recreate the directory structure on the new computer.
|
||||
<LI>Copy all necessary files across:
|
||||
<ul>
|
||||
<li> A copy of the SICServer.
|
||||
<li> A copy of the instrument configuration file.
|
||||
<li> All macro scripts used by this instrument.
|
||||
<li> The DataNumber file.
|
||||
<li> The necessary data file dictionary or header description files.
|
||||
<li> The client programs.
|
||||
</ul>
|
||||
All necessary files should be available from backup or under the
|
||||
/data/lnslib/sics/instrument tree on lnsa10.
|
||||
<LI>Edit the instrument configuration file and make all pathnames reflect the
|
||||
new directory structure. It should be enough to adapt the home variable at the
|
||||
top, but better check. All paths references must be renewed if the directory
|
||||
structure is changed from the one described <a href = "setup.htm">above</a>.
|
||||
<LI>Edit the client scripts to reflect the new location of the server. Only the
|
||||
box and logpath parameters should need to be changed.
|
||||
<LI> Edit the DataNumber file and set the number above the last valid data file
|
||||
number used. Otherwise the automatic numbering scheme
|
||||
for data files gets messed up.
|
||||
<LI>Test the new Installation!
|
||||
</OL>
|
||||
This requires the following steps:
|
||||
<ol>
|
||||
<li>Create a new local account on the host computer. There is a
|
||||
prefabricated account with the credentials: INSTBCK/INSTBCKLNS on
|
||||
lnsa15.
|
||||
<li>Run <b>sicsinstall <tt>instrument</tt> </b> in the new instruemnt
|
||||
account, thereby replacing instrument with the name of the instrument
|
||||
you are moving.
|
||||
<li>Create and edit a suitable DataNumber file for the instrument.
|
||||
<li>Edit the instrument configuration files and adapt the path names
|
||||
to match the new situation.
|
||||
<li>Configure the histogram memory to boot from the new computer, se
|
||||
histogram memory documsntation for instructions how to do that.
|
||||
<li>Try to start and debug.
|
||||
</ol>
|
||||
</P>
|
||||
<h2>Exchanging the Macintosh Serial Port Server</h2>
|
||||
<p>In its current configuration at SINQ, SICS requires a Macintosh PC running a serial
|
||||
port server program. Exchanging this Macintosh requires the following steps:
|
||||
<OL>
|
||||
<LI>Exchange the Macintosh physically. Plug everything as it was before the
|
||||
exchange. Not every Macintosh can be used. Any other Macintosh from another
|
||||
instrument or the spare Macintosh from the EDV-cabin will do. Other
|
||||
Macintoshs require some serious screwing and software installation
|
||||
procedures.
|
||||
<LI>Edit the instrument configuration file and replace all references to the name of
|
||||
the old Macintosh PC with the name of the new Macintosh PC.
|
||||
<LI>Test it!
|
||||
</OL>
|
||||
If you do not plug the serial devices back onto the Macintosh as they were, you
|
||||
need to adapt the channel numbers in the configuration file as well. The standard
|
||||
plugging scheme for the serial port extension box on the Macintosh is like this:
|
||||
<OL>
|
||||
<LI>Motor Controller 1
|
||||
<LI>Motor Controller 2
|
||||
<LI>Counter Box
|
||||
</OL>
|
||||
|
||||
<h2>Exchanging the Serial Port Server</h2>
|
||||
<p>
|
||||
<ol>
|
||||
<li>Fetch a new one and make sure that all cables are plugged as
|
||||
they were in the old one.
|
||||
<li>Edit the startsics script to start the SerPortServer program with
|
||||
the name of the new serial port server.
|
||||
<li>Done!
|
||||
</ol>
|
||||
</p>
|
||||
<h2>Exchanging the Histogram Memory</h2>
|
||||
<p>
|
||||
<ol>
|
||||
<li>Get a new histogram memory computer from either Peter Rasmussen,
|
||||
the test setup in WHGA/247 or in cases of greatest need, from SLS.
|
||||
<li>Put into the rack.
|
||||
<li>Configure the HM boot parameters through the console conneted to
|
||||
the serial port belonging to the HM. Instructions for this can be
|
||||
found in the histogram memory documentation.
|
||||
<li>Include the new HM into the /etc/hosts file of the instrument
|
||||
computer.
|
||||
<li>Adapt the instrument configuration file to reflect the new name of
|
||||
the HM.
|
||||
<li>Start and debug.
|
||||
</ol>
|
||||
</p>
|
||||
</BODY>
|
||||
</HTML>
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
Reference in New Issue