PSI update

r1464 | ffr | 2007-02-12 12:20:21 +1100 (Mon, 12 Feb 2007) | 2 lines
This commit is contained in:
Ferdi Franceschini
2007-02-12 12:20:21 +11:00
committed by Douglas Clowes
parent 634f2023b1
commit 3168325921
157 changed files with 29053 additions and 910 deletions

View File

@@ -29,6 +29,18 @@ maximum number of cycles was reached. This routine requires that the
instrument is currently placed somewhere on the peak and not miles away.
</P>
<p>
The peak optimiser supports another optimisation algorithm which is faster but
may not be as accurate. This is hill climbing:
<pre>
while errors gt precision and cycles lt maxcycles
for all variables
find the direction into which the intensity rises
step into this direction until the intensity drops
end for
end while
</pre>
</p>
<p>
The Peak Optimiser is implemented as an object with the name opti. It
understand the following commands:
<DL>
@@ -43,7 +55,10 @@ and number of steps parameters should cover the whole peak. However, the
Optimiser will extend the scan is the specified range is not sufficient.
<DT>opti run
<DD>Starts the optimiser. It will then optimise the peak. This may take some
time.
time as it uses a time consuming scan based algorithm.
<DT>opti climb
<DD>Starts the optimiser in hill climbing mode. Hill climbing is faster but may
not be as accurate as a scan based optimization.
</DL>
The behaviour of the optimiser can be configured by modifying some
parameters. The synatx is easy: <b>opti parameter</b> prints the value of the
@@ -72,5 +87,12 @@ status of the countmode parameter this is either a preset time or a preset
monitor.
</DL>
</p>
<p>
It is the users reponsability to provide meaningful step widths. Usually this is
dependent on the instrument resolution and thus fairly constant. Also these
optimisation algorithms will fail if the instrument is not positioned at the
flank of a peak. Probaly the best will be to do several cycles of hill
climbing first, followed by one cycle of scan optimisation for extra accuracy.
</p>
</BODY>