PSI UPDATE
r2720 | ffr | 2008-10-13 15:40:07 +1100 (Mon, 13 Oct 2008) | 2 lines
This commit is contained in:
@ -66,6 +66,10 @@ above and restores SICS to the state it was in when the status was saved with
|
||||
read.
|
||||
</p>
|
||||
<p>
|
||||
<b>restore listerr</b> prints the list of lines which caused errors during the
|
||||
last restore.
|
||||
</p>
|
||||
<p>
|
||||
<b>killfile</b> decrements the data number used for SICS file writing
|
||||
and thus consequently overwrites the last datafile. This is useful
|
||||
when useless data files have been created during tests. As this is
|
||||
|
@ -94,15 +94,23 @@ In order to calculate a UB matrix a list of reflections must be maintained. This
|
||||
This section covers the parameters and commands to use to make the module do calculations
|
||||
for you.
|
||||
<dl>
|
||||
<dt>tasbub const ki | kf
|
||||
<dt>tasbub const ki | kf | elastic
|
||||
<dd>Sets a parameter to determine if KI or KF is fixed when the energy transfer en is
|
||||
being driven. Allowed values: ki, kf
|
||||
being driven. Allowed values: ki, kf, elastic. In elastic mode the analyzer is
|
||||
disregarded. This is useful for two circle diffractometers.
|
||||
<dt>tasub const
|
||||
<dd>Prints if ki or kf is fixed.
|
||||
<dt>tasub ss
|
||||
<dd>Prints the sample scattering sense.
|
||||
<dt>tasub ss 1 | -1
|
||||
<dd>Sets the sample scattering sense. Allowed values are either 1 or -1.
|
||||
<dt>tasub silent 0 | 1
|
||||
<dd>Prints or sets the silent flag. If this is 0, the messages Driving motor ..
|
||||
from .. to .. are suppressed.</dd>
|
||||
<dt>tasub outofplane 0 | 1
|
||||
<dd>Prints or sets the outofplane flag. If this flag is 0, the instrument will stay
|
||||
in the scattering plane and not move out of it. This is here in order to protect those
|
||||
bloody magnets which cannot be tilted.</dd>
|
||||
<dt>tasub makeub r1 r2
|
||||
<dd>Calculate a new UB matrix from the current cell constants and the entries r1 and r2 in
|
||||
the reflection list. r1 and r2 are integer numbers. This command will not only print the
|
||||
@ -143,7 +151,7 @@ The virtual motor qm implements <b>powder mode</b>. In this mode, only the sampl
|
||||
respective positions. THis is commonly used to analyze the energy transfer of powder samples.
|
||||
</p>
|
||||
<p>
|
||||
There is another important command:
|
||||
There are other important command:
|
||||
<dl>
|
||||
<dt>tasub update
|
||||
<dd>This command will force a recalculation of the current Q-E position for the virtual
|
||||
@ -151,6 +159,9 @@ motors from angles. Normally tasub will take care of this. However, if any of th
|
||||
motors are moved directly or manualy, this command might be required. The SICS dr
|
||||
wrapper command, however, even takes care of this.
|
||||
</dl>
|
||||
<dt>tasub updatetargets
|
||||
<dd>This command makes the QE targets macth the current position. This is
|
||||
useful after initialization in the instrument.tcl file.</dd>
|
||||
</p>
|
||||
<h3>Internal Commands</h3>
|
||||
<p>
|
||||
|
Reference in New Issue
Block a user