deviceDesc = orange cryostat with 50 mm sample space for ULT
device stick_menu "heliox dil2 dil3 dil4 ori7"
makenv tt -driver [lsdriver] {
lsc_sensor tm B -sensor x174786 -sensorname VTI
lsc_sensor ts C -sensorname sample
lsc_sensor ts_2 D -sensor code -sensorname sample2
lsc_loop set -loop 1 -channel B -maxheater 50W -resist 50Ohm -color green
}
tt set/limit 310
makeCCU4 ori7 nv n2 he
hupdate /nv/set 0.6
nv ctrl/prop_o 0.05
nv ctrl/prop_c 0.03
nv ctrl/minpulse_c 0.005
nv ctrl/minpulse_o 0.01
nv ctrl/hystpulse_o 0.05
nv ctrl/hystpulse_c 0.05
nv ctrl/openpulse 5
nv ctrl/adjust_minpulse 0
makenv table -driver table /tt/tm {
"tt set/prop" "50:50 300:20"
"tt set/integ" "10:50 50:20 300:5"
"tt dblctrl/int2" "2:10 20:50 50:200 120:300 300:600"
"tt dblctrl/prop_up" "20:0.1 50:0.4"
"tt dblctrl/prop_lo" "20:0.1 50:0.8"
}
table tbl_tt_dblctrl_prop_lo 20:0.1 50:0.4
makenv dilpars -driver dilpars -condenseflow 5 -circulateflow 1.6
#GraphAdd cc.mp sec pulse red
#GraphAdd cc.msp sec speed green