deviceDesc = 30 K - 475 K closed cycle cryostat with hot stage
defineTemperature tt
makenv tt -driver [lsdriver] {
lsc_sensor tm B -sensor rhfe4140 -sensorname main
lsc_sensor te A -sensor x89009 -sensorname coldhead -alarm 300
lsc_loop set -loop 1 -channel B -maxheater 50W -resist 50Ohm
}
tt set/limit 470
tt set/prop 5
tt set/integ 2
tt tm/alarm 490
catch {
enableTable 0
}
appendVars tt.te/K/ColdFinger_T/6
# ambigous code for CTI7 and CTI7HT
#makeCCU4 cti7ht