deviceDesc = MA15 LN2 cooldown (does it work?) defineTemperature tt device stick_menu coil makenv tt -driver [lsdriver] { lsc_sensor tm A -sensor x28611 -sensorname VTI lsc_sensor ts C -sensorname sample lsc_sensor ts_2 D -sensor code -sensorname sample2 lsc_loop set -loop 1 -channel A -maxheater 80W -resist 40Ohm -color green if {[lsdriver] eq "336_lsc"} { lsc_loop setsamp -loop 2 -channel C -maxheater 10W -resist 300Ohm -color dark_green } } tt set/limit 310 makeCCU4 ma15 nv he_ilm catch { hupdate /nv/set 2.2 nv ctrl/minpulse_o 0.01 nv ctrl/minpulse_c 0.05 nv tol 0.02 } makeIps -limit 14.9 -startRamp 0.31 -port ma15-ts:3002 catch { mf limit 13.5 mf perscurrent 1.0 } makeIlm 2 -port ma15-ts:3001 makeN2Fill ccu4ilm { ln2fill lowlevel 30 ln2fill highlevel 90 ln2fill minfillminutes 8 ln2fill maxfillminutes 40 ln2fill minholdhours 24 ln2fill maxholdhours 96 # appendVars ln2fill.smooth/%/N2smooth/13 } # renamed ta...td, but kept the labels in order (M.Z. 27.8.2015) makenv tcoil -driver mom -port ma15-ts:3003 { mom_sensor td -label OnLambda -curve c_ab -adjust {223.0 295 242.5 77.4 1151 4.22 4772 2.16} mom_sensor tb -label AboveLambda -curve c_ab -adjust {222.0 295 241.14 77.4 1147 4.22 4882 2.16} mom_sensor tc -label MagTop -curve x30906 -adjust {194.0 295 362.78 77.4 3968 4.22 11180 2.16} mom_sensor ta -label MagBot -curve c_ab -adjust {170.5 295 192.33 77.4 1125 4.22 5064 2.16} } appendVars n2coolPar.istate|i|State|1 lc|mbar|Lambda_p|4 makeN2cool "tcoil ta" catch { # finish hour (tomorrow morning) # previous procedure: # day 1, morning: start N2, day2 morning: start fill (fast, He vessel pressurized) # new procedure # day 1, morning or noon: start N2:, day2 afternoon: slow fill overnight (self pressured He vessel) n2coolPar time 12 # target temperature for MagBot n2coolPar topLimit 85 # cooling rate K / min. n2coolPar coolRate 0.3 } GraphAdd n2coolpar.reg|K_2|T_reg|green GraphAdd n2coolpar.tfil|K_2|T_filtered|red hsetprop /LN2fill group "LN2 fill (not for LN2 precool)" makenv lnv -driver nv_ccu -controller _lnv -port ma15-ts:3004 "lambda needle valve"