deviceDesc = dry system if {"[sicsdescriptor t]" != "notfound"} { clientput [listexe] stopexe t clientput success success clientput "removeobject t" removeobject t } # --- configuration script config listen 1 # create the device # change X77456 for X154360 when calibration exists ## Configuration for attocube testing: #makenv tt -driver [lsdriver] { # lsc_sensor tm A -sensor X71741 -sensorname top # lsc_sensor ts B -sensor X68433 -sensorname bottom # lsc_sensor tc C -sensor X154360 -sensorname middle # lsc_loop set -loop 1 -channel B -maxheater 25W -resist 25Ohm #} # Configuration for braid thermal conductivity measurements: makenv tt -driver [lsdriver] { lsc_sensor ts B -sensor X154360 -sensorname stage2 lsc_loop set -loop 1 -channel B -maxheater 25W -loopname braid -resist 50Ohm } tt set/limit 315 tt set/prop 15 tt set/integ 10 catch { enableTable 0 } makenv tx -driver 370_lsc { lsc_sensor S1 -channel 1 -sensor ruox002 lsc_sensor S2 -channel 3 -sensor ruox001 #please set sample Thermpmeter to the real calib GraphAdd tx.S1 K S1 yellow GraphAdd tx.S2 K S2 cyan } # --- compressor --- makenv cp2800 cp2800 GraphAdd cp2800.inp_water_t C InpWater_C red GraphAdd cp2800.out_water_t C OutWater_C blue GraphAdd cp2800.oil_t C Oil_C green GraphAdd cp2800.helium_t C Helium_C magenta GraphAdd cp2800.high_side_p_min psi High_psi_min red GraphAdd cp2800.high_side_p_max psi High_psi_max orange GraphAdd cp2800.low_side_p_min psi Low_psi_min green GraphAdd cp2800.low_side_p_max psi Low_psi_max dark_green