stickDesc = dilution insert (added channels 2 blue, 9 yellow, 16 red) stick_sensors undefined undefined makenv ts -driver dil12 -reg370 /treg -mon370 /tmon -hasturbo 1 # lakeshore 370 ctrl makenv treg -driver 370_lsc -port dil4-ts:3001 { # curves for sample: # u02045 normal dil2 holder # m59 weak link 2 # ruoxben1k weak link with test ring heater # ruox5k weak link with test ring heater # standard ch6 cx078 (new sensor from Sep2014) # heater/thermo ch8 rx078 (new field insensitive sensor Sep2014) lsc_sensor sample -channel 6 -sensor cx078 -active 0 -excitation 20uV \ -color blue lsc_sensor mix -channel 5 -sensor ruoxm0 -active 1 -excitation 20uV \ -color cyan lsc_sensor samplehtr -channel 8 -sensor rx078 -active 0 -excitation 20uV \ -color black # weaklink heater lsc_loop set -channel 5 -maxheater 31mA -resist 316 -linearpower 0.2e-3 GraphAdd treg.set.reg K T_set dark_green GraphAdd treg.set.power W Htr_Power dark_green } # fix a bug (in the 370?): set the HRTRNG to 7 (31mA) when needed hsetprop /treg/set fix_range 7 # lakeshore 370 monitor makenv tmon -driver 370_lsc -port dil4-ts:3003 { lsc_sensor sorb -channel 12 -sensor c270 -excitation 2mV -color dark_violet lsc_sensor onek -channel 11 -sensor ruoxm0 -excitation 2mV -color yellow lsc_sensor stillt -channel 10 -sensor ruoxm0 -excitation 200uV -color orange lsc_sensor rblue -channel 2 -sensor raw -excitation 20uV lsc_sensor ryellow -channel 9 -sensor raw -excitation 20uV lsc_sensor rred -channel 16 -sensor raw -excitation 20uV lsc_sensor sample 3 clone lsc_sensor samplehtr 1 clone lsc_sensor mix 4 clone } proc graphvti {} { GraphItem shown tt.ts 0 GraphItem label tt.set.reg VTI_reg GraphItem label tt.set.power VTI_power } dolater 0 graphvti tt ts/curve undefined hsetprop /tt group "VTI temperature" makeDil -port dil4-ts:3005 makenv sorb -driver piloop -invar "/tmon/sorb" -outvar "dil psorb" \ -prop 200 -int 15 #makenv n2trap -driver n2_ccu -port dil4-ts:3007 "trap level" trap # control on mix by default ts control 6 treg set/deriv 0 defineTemperature ts dil extVersion 1 # take last selected values when doing samenv reload catch {default ts control} catch {default ts heaterselect} #makenv -objname condflow -driver bronkfreg -port dil4-ts:3008 200. condflow #GraphAdd condflow mln-min condflow magenta