87c7a791f5
issue with He auto fill when external levelmeter is used
...
needed also a fix in CCU4 firmware (new Version 8.3)
2023-09-12 10:46:49 +02:00
d11d9db560
fix valve readback
...
the ccu4 firmware does not report changes on valves upon the ?
command. fix it: poll explicitly
2023-08-08 14:28:26 +02:00
l_samenv
366ad2d560
fix hfflow config
...
+ ccu4: make unused valves invisible
2023-05-24 11:48:03 +02:00
e91a753b87
Merge branch 'master' of gitlab.psi.ch-samenv:samenv/sea
2022-10-24 14:46:21 +02:00
fd176cd38c
fix he level edge cases
...
- a full He reservoir might lead to a timeout message.
a value of 111 is used instead of a "timeout" error
- likewise, on a warm He reservoir -11.1 is used instead of
the sensor warm message
2022-10-24 14:42:26 +02:00
l_samenv
6d99df5bf1
CCU4: fix enum in he ext channel parameter
2022-10-17 17:09:43 +02:00
l_samenv
46dcf8e77c
initial commit