893bd5fa4b
automatic LN2 fill with ILM - fix problems on restart
...
on the first reading, the level value might be invalid,
then it tries to fill and switch to inactive because there
is not valid level value.
fix: stay watching as long as the level is invalid
2025-07-04 10:51:54 +02:00
33296cc03b
do not trigger error on level reading when inactive
2023-10-05 08:24:52 +02:00
l_samenv
3773bb98b3
fix autofill with extrnal level meter
...
when no value is available, filling must be inhibited.
this patch checks whether a value newer than 30 sec is
available
2023-09-15 10:37:13 +02:00
87c7a791f5
issue with He auto fill when external levelmeter is used
...
needed also a fix in CCU4 firmware (new Version 8.3)
2023-09-12 10:46:49 +02:00
911fdd3106
remove fill debug message
2023-09-01 09:31:03 +02:00
l_samenv
969bfe95f1
add manualfill to ccu4ext
2023-03-31 14:34:37 +02:00
l_samenv
46dcf8e77c
initial commit
2022-08-18 15:04:28 +02:00