From b5540c9dd66bf7ef520756359c34ec537bce83dd Mon Sep 17 00:00:00 2001 From: camea Date: Thu, 6 Jun 2024 09:06:29 +0200 Subject: [PATCH] add special config files for PE cell --- tcl/ccrpe_lowT.config | 33 +++++++++++++++++++++++++++++++++ tcl/flukepe.addon | 6 ++++++ 2 files changed, 39 insertions(+) create mode 100644 tcl/ccrpe_lowT.config create mode 100644 tcl/flukepe.addon diff --git a/tcl/ccrpe_lowT.config b/tcl/ccrpe_lowT.config new file mode 100644 index 0000000..1f1efd2 --- /dev/null +++ b/tcl/ccrpe_lowT.config @@ -0,0 +1,33 @@ +deviceDesc = 4 K closed cycle cryostat (PE cell) + +defineTemperature tt +makenv tt -driver [lsdriver] { + lsc_sensor tm A -sensor x33924 -sensorname main + lsc_sensor ts C -sensor x31320 -sensorname sample + lsc_sensor te D -sensor psam -sensorname screen + lsc_loop set -loop 1 -channel A -maxheater 100W -resist 25Ohm + if {[lsdriver] eq "336_lsc"} { + lsc_loop current -loop 3 -channel C -maxheater 0.59W -resist 220Ohm -loopname pewarmup -color hidden + } +} +tt set/limit 320 + +appendVars tt.te/K/firstStage_T/7 +catch { + enableTable 0 +} + +#makeN2Fill + +#makenv ep euro2k +#obj_list ep EuroThermPressure +#appendVars ep/%/Pressure + +makeCCU4 ccrpe nv + +makenv lev -driver ami136 + +makenv warmup -driver piloop -invar "tt" -outvar "tt current/manualpower" \ + -prop 0.0005 -int 120 -outmin 0 -outmax 0.053 + +GraphAdd warmup.output kW Htr_wup auto diff --git a/tcl/flukepe.addon b/tcl/flukepe.addon new file mode 100644 index 0000000..7a545d2 --- /dev/null +++ b/tcl/flukepe.addon @@ -0,0 +1,6 @@ +addonDesc = pe pressure over fluke 45 multimeter + +makenv p -driver fluke45 {X * 500 * 900 + 0.0} + +GraphAdd p bar p dark_green +