frappy/cfg/sim_mlz_ccidu1_cfg.py
Alexander Zaft 8799710b38 Convert example configs to python
Fixes: #4627

Change-Id: I1049af9fa2f3f9ec06e55811dc9ecfa5f359c805
2023-03-06 08:24:15 +01:00

107 lines
3.2 KiB
Python

Node('ccidu1',
'[sim] ccidu box of MLZ Sample environment group\n'
'\n'
'Controls an 3He/4He dilution insert with an ls372 and an PLC controlling\n'
'the compressor and the valves of the gas handling.',
'tcp://10767',
)
Mod('T_ccidu1',
'frappy.simulation.SimDrivable',
'Main temperature control node of ccidu1.\n'
'\n'
'Controls the regulation loop of the ls372.',
value = Param(default=300, unit='K'),
target = Param(default=300, datatype={"type":"double", "min":0, "max":300, "unit":"K"}),
extra_params='ramp',
ramp = Param(default=60,
datatype={"type":"double", "min":0, "max":600, "unit":"K/min"},
description='target ramping speed in K/min.',
),
meaning=["temperature_regulation",40]
)
Mod('T_ccidu1_A',
'frappy.simulation.SimReadable',
'mixing chamber temperature sensor. Also used for the regulation.',
value = Param(default=300, datatype={"type":"double", "unit":"K"}),
visibility='expert',
meaning=["temperature",38],
)
Mod('T_ccidu1_B',
'frappy.simulation.SimReadable',
'(optional) sample temperature sensor close to sample.',
value = Param(default=300, datatype={"type":"double", "unit":"K"}),
visibility='user',
meaning=["temperature",39],
)
Mod('ccidu1_pstill',
'frappy.simulation.SimReadable',
'Pressure at the still/turbo pump inlet.',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_pinlet',
'frappy.simulation.SimReadable',
'Pressure at forepump inlet/turbo pump outlet.',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_poutlet',
'frappy.simulation.SimReadable',
'Pressure at forepump outlet/compressor inlet.',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_pkond',
'frappy.simulation.SimReadable',
'Pressure at condensing line/compressor outlet.',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_ptank',
'frappy.simulation.SimReadable',
'Pressure in dump tank.',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_pvac',
'frappy.simulation.SimReadable',
'Pressure in the vacuum dewar (ivc).',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
Mod('ccidu1_flow',
'frappy.simulation.SimReadable',
'Gas Flow (condensing line).',
value = Param(default=999, datatype={"type":"double", "unit":"mbar"}),
visibility='expert',
)
# note: all valves and switches are missing: use VNC to control them
Mod('ccidu1_V6',
'frappy.simulation.SimDrivable',
'Needle valve',
value = Param(default=99, datatype={"type":"double", "min":0, "max":100, "unit":"%%"}),
visibility='expert',
)
Mod('ccidu1_V3',
'frappy.simulation.SimWritable',
'Dump Valve',
value = Param(
default="OFF",
datatype={"type":"enum", "members":{"on": 1, "OFF":0}}
),
target = Param(datatype={"type":"enum", "members":{"on": 1, "OFF":0}}),
visibility='expert',
)