# This program is free software; you can redistribute it and/or modify it under
|
|
# the terms of the GNU General Public License as published by the Free Software
|
|
# Foundation; either version 2 of the License, or (at your option) any later
|
|
# version.
|
|
#
|
|
# This program is distributed in the hope that it will be useful, but WITHOUT
|
|
# ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS
|
|
# FOR A PARTICULAR PURPOSE. See the GNU General Public License for more
|
|
# details.
|
|
#
|
|
# You should have received a copy of the GNU General Public License along with
|
|
# this program; if not, write to the Free Software Foundation, Inc.,
|
|
# 59 Temple Place, Suite 330, Boston, MA 02111-1307 USA
|
|
#
|
|
# Module authors:
|
|
# Oksana Shliakhtun <oksana.shliakhtun@psi.ch>
|
|
# Markus Zolliker <markus.zolliker@psi.ch>
|
|
# *****************************************************************************
|
|
"""bath thermostat Thermo Scientific™ ARCTIC A10 Refrigerated Circulators"""
|
|
|
|
from frappy.core import Command, StringIO, Parameter, HasIO, \
|
|
Drivable, FloatRange, IDLE, BUSY, ERROR, WARN, BoolType
|
|
from frappy.structparam import StructParam
|
|
from frappy_psi.convergence import HasConvergence
|
|
|
|
|
|
class ThermFishIO(StringIO):
|
|
end_of_line = '\r'
|
|
identification = [('RVER', r'.*')] # Firmware Version
|
|
|
|
|
|
class TemperatureLoopA10(HasConvergence, HasIO, Drivable):
|
|
ioClass = ThermFishIO
|
|
value = Parameter('temperature', unit='degC')
|
|
target = Parameter('setpoint/target', datatype=FloatRange, unit='degC', default=0)
|
|
control_active = Parameter('circilation and control is on', BoolType(), default=False)
|
|
ctrlpars = StructParam('control parameters struct', dict(
|
|
p_heat = Parameter('proportional heat parameter', FloatRange()),
|
|
i_heat = Parameter('integral heat parameter', FloatRange()),
|
|
d_heat = Parameter('derivative heat parameter', FloatRange()),
|
|
p_cool = Parameter('proportional cool parameter', FloatRange()),
|
|
i_cool = Parameter('integral cool parameter', FloatRange()),
|
|
d_cool = Parameter('derivative cool parameter', FloatRange()),
|
|
), readonly=False)
|
|
|
|
status_messages = [
|
|
(ERROR, 'high tempr. cutout fault', 2, 0),
|
|
(ERROR, 'high RA tempr. fault', 2, 1),
|
|
(ERROR, 'high temperature fixed fault', 3, 7),
|
|
(ERROR, 'low temperature fixed fault', 3, 6),
|
|
(ERROR, 'high temperature fault', 3, 5),
|
|
(ERROR, 'low temperature fault', 3, 4),
|
|
(ERROR, 'low level fault', 3, 3),
|
|
(ERROR, 'circulator fault', 4, 5),
|
|
(ERROR, 'high press. cutout', 5, 2),
|
|
(ERROR, 'motor overloaded', 5, 1),
|
|
(ERROR, 'pump speed fault', 5, 0),
|
|
(WARN, 'open internal sensor', 1, 7),
|
|
(WARN, 'shorted internal sensor', 1, 6),
|
|
(WARN, 'high temperature warn', 3, 2),
|
|
(WARN, 'low temperature warn', 3, 1),
|
|
(WARN, 'low level warn', 3, 0),
|
|
(IDLE, 'max. heating', 5, 5),
|
|
(IDLE, 'heating', 5, 6),
|
|
(IDLE, 'cooling', 5, 4),
|
|
(IDLE, 'max cooling', 5, 3),
|
|
(IDLE, '', 4, 3),
|
|
]
|
|
|
|
def get_par(self, cmd):
|
|
"""get parameter and convert to float
|
|
|
|
:param cmd: hardware command without the leading 'R'
|
|
|
|
:return: result converted to float
|
|
"""
|
|
new_cmd = 'R' + cmd
|
|
reply = self.communicate(new_cmd).strip()
|
|
while reply[-1].isalpha():
|
|
reply = reply[:-1]
|
|
return float(reply)
|
|
|
|
def set_par(self, cmd, value):
|
|
self.communicate(f'S{cmd} {value}')
|
|
return self.get_par(cmd)
|
|
|
|
def read_value(self):
|
|
"""
|
|
Reading internal temperature sensor value.
|
|
"""
|
|
return self.get_par('T')
|
|
|
|
def read_status(self):
|
|
""" convert from RUFS Command: Description of Bits
|
|
|
|
====== ======================================================== ===============
|
|
Value Description
|
|
====== ======================================================== ===============
|
|
V1 B6: warning, rtd1 (internal temp. sensor) is shorted B0 --> 1
|
|
B7: warning, rtd1 is open B1 --> 2
|
|
V2 B0: error, HTC (high temperature cutout) fault B2 --> 4
|
|
B1: error, high RA (refrigeration) temperature fault B3 --> 8
|
|
V3 B0: warning, low level in the bath B5 --> 32
|
|
B1: warning, low temperature B6 --> 64
|
|
B2: warning, high temperature B7 --> 128
|
|
B3: error, low level in the bath
|
|
B4: error, low temperature fault
|
|
B5: error, high temperature fault
|
|
B6: error, low temperature fixed* fault
|
|
B7: error, high temperature fixed** fault
|
|
V4 B3: idle, circulator** is running
|
|
B5: error, circulator** fault
|
|
V5 B0: error, pump speed fault
|
|
B1: error, motor overloaded
|
|
B2: error, high pressure cutout
|
|
B3: idle, maximum cooling
|
|
B4: idle, cooling
|
|
B5: idle, maximum heating
|
|
B6: idle, heating
|
|
====== ======================================================== ===============
|
|
"""
|
|
result_str = self.communicate('RUFS') # read unit fault status
|
|
values_str = result_str.strip().split()
|
|
values_int = [int(val) for val in values_str]
|
|
|
|
for status_type, status_msg, vi, bit in self.status_messages:
|
|
if values_int[vi-1] & (1 << bit):
|
|
conv_status = HasConvergence.read_status(self)
|
|
if self.isBusy(conv_status):
|
|
# use 'inside tolerance' and 'outside tolerance' from HasConvergence,
|
|
# else our own status
|
|
return BUSY, conv_status[1] if 'tolerance' in conv_status[1] else status_msg
|
|
return status_type, status_msg
|
|
return WARN, 'circulation off'
|
|
|
|
def read_control_active(self):
|
|
return int(self.get_par('O'))
|
|
|
|
@Command
|
|
def control_off(self):
|
|
"""switch control and circulation off"""
|
|
self.control_active = self.set_par('O', 0)
|
|
|
|
def read_target(self):
|
|
return self.get_par('S')
|
|
|
|
def write_target(self, target):
|
|
self.control_active = self.set_par('O', 1)
|
|
self.communicate(f'SS {target}')
|
|
self.convergence_start()
|
|
return target
|
|
|
|
def read_p_heat(self):
|
|
return self.get_par('PH')
|
|
|
|
def write_p_heat(self, value):
|
|
return self.set_par('PH', value)
|
|
|
|
def read_i_heat(self):
|
|
return self.get_par('IH')
|
|
|
|
def write_i_heat(self, value):
|
|
return self.set_par('IH', value)
|
|
|
|
def read_d_heat(self):
|
|
return self.get_par('DH')
|
|
|
|
def write_d_heat(self, value):
|
|
return self.set_par('DH', value)
|
|
|
|
def read_p_cool(self):
|
|
return self.get_par('PC')
|
|
|
|
def write_p_cool(self, value):
|
|
return self.set_par('PC', value)
|
|
|
|
def read_i_cool(self):
|
|
return self.get_par('IC')
|
|
|
|
def write_i_cool(self, value):
|
|
return self.set_par('IC', value)
|
|
|
|
def read_d_cool(self):
|
|
return self.get_par('DC')
|
|
|
|
def write_d_cool(self, value):
|
|
return self.set_par('DC', value)
|