frappy/secop_psi/k2601b.py
Markus Zolliker 07b758c3dd removed old style syntax
- removed secop/metaclass.py
- moved code from ModuleMeta to modules.HasAccessibles.__init_subclass__
- reworked properties:
  assignment obj.property = value now always allowed
- reworked Parameters and Command to be true descriptors
- Command must now be solely used as decorator
- renamed 'usercommand' to 'Command'
- command methods no longer start with 'do_'
- reworked mechanism to determine accessible order:
  the attribute paramOrder, if given, determines order of accessibles
+ fixed some issues makeing the IDE more happy
+ simplified code for StatusType and added a test for it

Change-Id: I8045cf38ee6f4d4862428272df0b12a7c8abaca7
Reviewed-on: https://forge.frm2.tum.de/review/c/sine2020/secop/playground/+/25049
Tested-by: Jenkins Automated Tests <pedersen+jenkins@frm2.tum.de>
Reviewed-by: Enrico Faulhaber <enrico.faulhaber@frm2.tum.de>
Reviewed-by: Markus Zolliker <markus.zolliker@psi.ch>
2021-02-26 17:27:13 +01:00

160 lines
5.6 KiB
Python

#!/usr/bin/env python
# -*- coding: utf-8 -*-
# *****************************************************************************
# This program is free software; you can redistribute it and/or modify it under
# the terms of the GNU General Public License as published by the Free Software
# Foundation; either version 2 of the License, or (at your option) any later
# version.
#
# This program is distributed in the hope that it will be useful, but WITHOUT
# ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS
# FOR A PARTICULAR PURPOSE. See the GNU General Public License for more
# details.
#
# You should have received a copy of the GNU General Public License along with
# this program; if not, write to the Free Software Foundation, Inc.,
# 59 Temple Place, Suite 330, Boston, MA 02111-1307 USA
#
# Module authors:
# Markus Zolliker <markus.zolliker@psi.ch>
# *****************************************************************************
"""Keithley 2601B source meter
not tested yet"""
from secop.core import Writable, Module, Parameter, Attached,\
BoolType, FloatRange, EnumType, HasIodev, StringIO
class K2601bIO(StringIO):
identification = [('print(localnode.description)', 'Keithley Instruments SMU 2601B.*')]
SOURCECMDS = {
1: 'reset()'
'smua.source.func = smua.OUTPUT_DCVOLTS '
'display.smua.measure.func = display.MEASURE_DCAMP '
'smua.source.autorangev = 1',
2: 'reset()'
'smua.source.func = smua.OUTPUT_DCAMPS '
'smua.source.autorangei = 1',
}
class SourceMeter(HasIodev, Module):
resistivity = Parameter('readback resistivity', FloatRange(unit='Ohm'), poll=True)
power = Parameter('readback power', FloatRange(unit='W'), poll=True)
mode = Parameter('measurement mode', EnumType(off=0, current=1, voltage=2),
readonly=False, default=0)
active = Parameter('output enable', BoolType(), readonly=False, poll=True)
iodevClass = K2601bIO
def read_resistivity(self):
return self.sendRecv('print(smua.measure.r())')
def read_power(self):
return self.sendRecv('print(smua.measure.p())')
def read_active(self):
return self.sendRecv('print(smua.source.output)')
def write_active(self, value):
return self.sendRecv('smua.source.output = %d print(smua.source.output)' % value)
# for now, mode will not be read from hardware
def write_mode(self, value):
if value == 0:
self.write_active(0)
else:
self.sendRecv(SOURCECMDS[value] + ' print(0)')
return value
class Current(HasIodev, Writable):
sourcemeter = Attached()
value = Parameter('measured current', FloatRange(unit='A'), poll=True)
target = Parameter('set current', FloatRange(unit='A'), poll=True)
active = Parameter('current is controlled', BoolType(), default=False) # polled from Current/Voltage
limit = Parameter('current limit', FloatRange(0, 2.0, unit='A'), default=2, poll=True)
def read_value(self):
return self.sendRecv('print(smua.measure.i())')
def read_target(self):
return self.sendRecv('print(smua.source.leveli)')
def write_target(self, value):
if not self.active:
raise ValueError('current source is disabled')
if value > self.limit:
raise ValueError('current exceeds limit')
return self.sendRecv('smua.source.leveli = %g print(smua.source.leveli)' % value)
def read_limit(self):
if self.active:
return self.limit
return self.sendRecv('print(smua.source.limiti)')
def write_limit(self, value):
if self.active:
return value
return self.sendRecv('smua.source.limiti = %g print(smua.source.limiti)' % value)
def read_active(self):
return self._sourcemeter.mode == 1 and self._sourcemeter.read_active()
def write_active(self, value):
if self._sourcemeter.mode != 1:
if value:
self._sourcemeter.write_mode(1) # switch to current
else:
return 0
return self._sourcemeter.write_active(value)
class Voltage(HasIodev, Writable):
sourcemeter = Attached()
value = Parameter('measured voltage', FloatRange(unit='V'), poll=True)
target = Parameter('set voltage', FloatRange(unit='V'), poll=True)
active = Parameter('voltage is controlled', BoolType(), poll=True)
limit = Parameter('current limit', FloatRange(0, 2.0, unit='V'), default=2, poll=True)
def read_value(self):
return self.sendRecv('print(smua.measure.v())')
def read_target(self):
return self.sendRecv('print(smua.source.levelv)')
def write_target(self, value):
if not self.active:
raise ValueError('voltage source is disabled')
if value > self.limit:
raise ValueError('voltage exceeds limit')
return self.sendRecv('smua.source.levelv = %g print(smua.source.levelv)' % value)
def read_limit(self):
if self.active:
return self.limit
return self.sendRecv('print(smua.source.limitv)')
def write_limit(self, value):
if self.active:
return value
return self.sendRecv('smua.source.limitv = %g print(smua.source.limitv)' % value)
def read_active(self):
return self._sourcemeter.mode == 2 and self._sourcemeter.read_active()
def write_active(self, value):
if self._sourcemeter.mode != 2:
if value:
self._sourcemeter.write_mode(2) # switch to voltage
else:
return 0
return self._sourcemeter.write_active(value)