Node('ccr12', '''[sim] CCR12 box of MLZ Sample environment group Contains a Lakeshore 336 and an PLC controlling the compressor and some valves. This is how we use it now.''', interface='tcp://10767', ) Mod('T_ccr12', 'frappy.simulation.SimDrivable', description='Main temperature control node of CCR12.\n\nSwitches between regulation on stick and regulation on tube depending on temperature requested.\nMay also pump gas for higher temperatures, if configured.\n\nNote: in nicos this is handled by its own class which manages T_ccr12_stick and T_ccr12_tube.\nin this simulation this module is isolated.', extra_params=['ramp'], ramp=Param( datatype=FloatRange(unit='K/min', minval=0.0, maxval=60.0), default=60.0, ), value=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), target=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), meaning=('temperature_regulation', 20), ) Mod('T_ccr12_stick', 'frappy.simulation.SimDrivable', description='Temperature regulation for the sample stick in ccr12.', extra_params=['ramp'], ramp=Param( datatype=FloatRange(unit='K/min', minval=0.0, maxval=60.0), default=60.0, ), value=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), target=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), meaning=('temperature_regulation', 15), ) Mod('T_ccr12_tube', 'frappy.simulation.SimDrivable', description='Temperature regulation for the tube of ccr12.', extra_params=['ramp'], ramp=Param( datatype=FloatRange(unit='K/min', minval=0.0, maxval=60.0), default=60.0, ), value=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), target=Param( datatype=FloatRange(unit='K', minval=0.0, maxval=600.0), default=300.0, ), meaning=('temperature_regulation', 10), ) Mod('T_ccr12_A', 'frappy.simulation.SimReadable', description='(optional) Sample temperature sensor.', value=Param( datatype=FloatRange(unit='K'), default=300.0, ), meaning=('temperature', 9), ) Mod('T_ccr12_B', 'frappy.simulation.SimReadable', description='(regulation) temperature sensor on stick.', value=Param( datatype=FloatRange(unit='K'), default=300.0, ), meaning=('temperature', 10), ) Mod('T_ccr12_C', 'frappy.simulation.SimReadable', description='Temperature at the coldhead.', value=Param( datatype=FloatRange(unit='K'), default=70.0, ), meaning=('temperature', 1), ) Mod('T_ccr12_D', 'frappy.simulation.SimReadable', description='(regulation) temperature at coupling to stick.', value=Param( datatype=FloatRange(unit='K'), default=80.0, ), meaning=('temperature', 2), ) Mod('ccr12_pressure_regulate', 'frappy.simulation.SimWritable', description='Selects on which Sensor the pressure regulation works, or switches it off.', visibility='expert', value=Param( datatype=EnumType('', off=0, p1=1, p2=2), default='off', ), target=Param( datatype=EnumType('', off=0, p1=1, p2=2), default='off', ), ) Mod('ccr12_compressor', 'frappy.simulation.SimDrivable', description='Switches the compressor for the cooling stage on or off.\n\nNote: This should always be on, except for fast heatup for sample change.', value=Param( datatype=EnumType('', off=0, on=1), default='on', ), target=Param( datatype=EnumType('', off=0, on=1), default='on', ), ) Mod('ccr12_gas_switch', 'frappy.simulation.SimWritable', description='Switches the gas inlet on or off.\n\nnote: in reality this switches itself off after 15min.\nnote: in reality this is interlocked with ccr12_vacuum_switch, only one can be on!\nnote: in this simulation this module is isolated.', value=Param( datatype=EnumType('', off=0, on=1), default='off', ), target=Param( datatype=EnumType('', off=0, on=1), default='off', ), ) Mod('ccr12_vacuum_switch', 'frappy.simulation.SimWritable', description='Switches the vacuum pumping valve on or off.\n\nnote: in reality this is interlocked with ccr12_gas_switch, only one can be on!\nnote: in this simulation this module is isolated.', value=Param( datatype=EnumType('', off=0, on=1), default='off', ), target=Param( datatype=EnumType('', off=0, on=1), default='off', ), ) Mod('ccr12_p1', 'frappy.simulation.SimReadable', description='Default pressure Sensor, linear scale 0..1000mbar', value=Param( datatype=FloatRange(unit='mbar'), default=999.0, ), ) Mod('ccr12_p2', 'frappy.simulation.SimReadable', description='Auxillary pressure Sensor.', value=Param( datatype=FloatRange(unit='mbar'), default=1e-06, ), ) Mod('ccr12_curve_p2', 'frappy.simulation.SimWritable', description='Curve for Aux pressure Sensor p2', visibility='expert', value=Param( default='TTR100', datatype=EnumType('', 0..10V=0, default=1, 0..9V=2, DI200=3, DI2000=4, TTR100=7, PTR90=8, PTR225/PTR237=9, ITR90=10, ITR100 curve D=11, ITR100 curve 2=12, ITR100 curve 3=13, ITR100 curve 4=14, ITR100 curve 5=15, ITR100 curve 6=16, ITR100 curve 7=17, ITR100 curve 8=18, ITR100 curve 9=19, ITR100 curve A=20, CMR361=21, CMR362=22, CMR363=23, CMR364=24, CMR365=25), ), target=Param( datatype=EnumType('', 0..10V=0, default=1, 0..9V=2, DI200=3, DI2000=4, TTR100=7, PTR90=8, PTR225/PTR237=9, ITR90=10, ITR100 curve D=11, ITR100 curve 2=12, ITR100 curve 3=13, ITR100 curve 4=14, ITR100 curve 5=15, ITR100 curve 6=16, ITR100 curve 7=17, ITR100 curve 8=18, ITR100 curve 9=19, ITR100 curve A=20, CMR361=21, CMR362=22, CMR363=23, CMR364=24, CMR365=25), ), ) Mod('ccr12_p1_limits', 'frappy.simulation.SimWritable', description='Limits for pressure regulation in P1.', visibility='expert', value=Param( datatype=TupleOf(FloatRange(unit='mbar', minval=0.0, maxval=1000.0), FloatRange(unit='mbar', minval=0.0, maxval=1000.0)), default=(0.0, 10.0), ), target=Param( datatype=TupleOf(FloatRange(unit='mbar', minval=0.0, maxval=1000.0), FloatRange(unit='mbar', minval=0.0, maxval=1000.0)), default=(0.0, 10.0), ), ) Mod('ccr12_p2_limits', 'frappy.simulation.SimWritable', description='Limits for pressure regulation in P2.', visibility='expert', value=Param( datatype=TupleOf(FloatRange(unit='mbar', minval=0.0, maxval=1000.0), FloatRange(unit='mbar', minval=0.0, maxval=1000.0)), default=(1e-05, 0.001), ), target=Param( datatype=TupleOf(FloatRange(unit='mbar', minval=0.0, maxval=1000.0), FloatRange(unit='mbar', minval=0.0, maxval=1000.0)), default=(1e-05, 0.001), ), )