[node stressihtf2_v2] description = [sim] Stressihtf2 box of MLZ Sample environment group . Controls an High Temperature Furnace with an Eurotherm and an PLC controlling some valves and checking cooling water. [interface tcp] type=tcp bindto=0.0.0.0 bindport=10767 [module T] class=secop.simulation.SimDrivable .description=Main temperature control node of Stressihtf2. value.datatype={"type":"double", "min":0, "unit":"degC"} value.default=20 target.datatype={"type":"double", "min":0, "max":2000, "unit":"degC"} target.default=20 .extra_params=ramp,regulationmode,abslimits,userlimits ramp.datatype={"type":"double", "min":0, "max":600, "unit":"K/min"} ramp.description=target ramping speed in K/min. ramp.default=60 abslimits.datatype={"type":"limit", "members":{"type":"double", "min":0, "max":2000, "unit":"degC"}} abslimits.default=[0,2000] abslimits.description=currently active absolute limits for the setpoint. depend on the regulationmode parameter (both/stick->0..600, tube->0..300K). userlimits.datatype={"type":"limit", "members":{"type":"double", "min":0, "max":2000, "unit":"degC"}} userlimits.default=[0,300] userlimits.description=current user set limits for the setpoint. must be inside abslimits. userlimits.readonly=False .meaning=['temperature_regulation', 10] [module T_sample] class=secop.simulation.SimReadable .description=(optional) Sample temperature sensor. .visibility=expert value.default=300 value.datatype={"type":"double", "min":0, "unit":"degC"} .meaning=["temperature", 9] [module N2] class=secop.simulation.SimWritable .description=Switches the N2 gas inlet on or off. .visibility=expert value.default='off' value.datatype={"type":"enum", "members":{'off':0,'on':1}} target.datatype={"type":"enum", "members":{'off':0,'on':1}} target.default='off' [module He] class=secop.simulation.SimWritable .description=Switches the He gas inlet on or off. .visibility=expert value.default='off' value.datatype={"type":"enum", "members":{'off':0,'on':1}} target.datatype={"type":"enum", "members":{'off':0,'on':1}} target.default='off' [module lamps] class=secop.simulation.SimWritable .description=Switches the heating lamps on or off. .visibility=expert value.default='on' value.datatype={"type":"enum", "members":{'off':0,'on':1}} target.datatype={"type":"enum", "members":{'off':0,'on':1}} target.default='on' [module water_ok] class=secop.simulation.SimReadable .description=Readout of the cooling water state. .visibility=expert value.default='ok' value.datatype={"type":"enum", "members":{'failed':0,'ok':1}}