frappy_psi.thermofisher improvements
- merge Loop with Sensor - make convergence work Change-Id: Iba0cafc524ada6d490b7a5c30f4127e77fd163f3
This commit is contained in:
parent
9e52665b88
commit
720d010c47
@ -14,10 +14,6 @@ Mod('T',
|
||||
'holder temperature',
|
||||
io='io',
|
||||
target=Param(max=100),
|
||||
)
|
||||
|
||||
Mod('Th',
|
||||
'frappy_psi.thermofisher.SensorA10',
|
||||
'heat exch. temperature',
|
||||
io='io',
|
||||
tolerance=0.5,
|
||||
settling_time=20,
|
||||
)
|
||||
|
@ -73,8 +73,8 @@
|
||||
|
||||
"""
|
||||
|
||||
from frappy.core import StringIO, Parameter, Readable, HasIO, \
|
||||
Drivable, FloatRange, IDLE, ERROR, WARN, BoolType
|
||||
from frappy.core import StringIO, Parameter, HasIO, \
|
||||
Drivable, FloatRange, IDLE, BUSY, ERROR, WARN, BoolType
|
||||
from frappy_psi.convergence import HasConvergence
|
||||
|
||||
|
||||
@ -83,9 +83,20 @@ class ThermFishIO(StringIO):
|
||||
identification = [('RVER', r'.*')] # Firmware Version
|
||||
|
||||
|
||||
class SensorA10(HasIO, Readable):
|
||||
class TemperatureLoopA10(HasConvergence, HasIO, Drivable):
|
||||
ioClass = ThermFishIO
|
||||
value = Parameter('internal temperature', unit='degC')
|
||||
value = Parameter('temperature', unit='degC')
|
||||
target = Parameter('setpoint/target', datatype=FloatRange, unit='degC', default=0)
|
||||
circ_on = Parameter('is circulation running', BoolType(), readonly=False, default=False)
|
||||
# pids
|
||||
p_heat = Parameter('proportional heat parameter', FloatRange(), readonly=False)
|
||||
i_heat = Parameter('integral heat parameter', FloatRange(), readonly=False)
|
||||
d_heat = Parameter('derivative heat parameter', FloatRange(), readonly=False)
|
||||
p_cool = Parameter('proportional cool parameter', FloatRange(), readonly=False)
|
||||
i_cool = Parameter('integral cool parameter', FloatRange(), readonly=False)
|
||||
d_cool = Parameter('derivative cool parameter', FloatRange(), readonly=False)
|
||||
|
||||
status_messages = [
|
||||
(ERROR, 'high tempr. cutout fault', 2, 0),
|
||||
(ERROR, 'high RA tempr. fault', 2, 1),
|
||||
@ -138,22 +149,14 @@ class SensorA10(HasIO, Readable):
|
||||
|
||||
for status_type, status_msg, vi, bit in self.status_messages:
|
||||
if values_int[vi-1] & (1 << bit):
|
||||
conv_status = HasConvergence.read_status(self)
|
||||
if self.isBusy(conv_status):
|
||||
# use 'inside tolerance' and 'outside tolerance' from HasConvergence,
|
||||
# else our own status
|
||||
return BUSY, conv_status[1] if 'tolerance' in conv_status[1] else status_msg
|
||||
return status_type, status_msg
|
||||
return WARN, 'circulation off'
|
||||
|
||||
|
||||
class TemperatureLoopA10(SensorA10, Drivable):
|
||||
value = Parameter('temperature', unit='degC')
|
||||
target = Parameter('setpoint/target', datatype=FloatRange, unit='degC', default=0)
|
||||
circ_on = Parameter('is circulation running', BoolType(), readonly=False, default=False)
|
||||
# pids
|
||||
p_heat = Parameter('proportional heat parameter', FloatRange(), readonly=False)
|
||||
i_heat = Parameter('integral heat parameter', FloatRange(), readonly=False)
|
||||
d_heat = Parameter('derivative heat parameter', FloatRange(), readonly=False)
|
||||
p_cool = Parameter('proportional cool parameter', FloatRange(), readonly=False)
|
||||
i_cool = Parameter('integral cool parameter', FloatRange(), readonly=False)
|
||||
d_cool = Parameter('derivative cool parameter', FloatRange(), readonly=False)
|
||||
|
||||
def read_circ_on(self):
|
||||
return self.communicate('RO')
|
||||
|
||||
@ -171,7 +174,7 @@ class TemperatureLoopA10(SensorA10, Drivable):
|
||||
"""
|
||||
self.write_circ_on('1')
|
||||
self.communicate(f'SS {target}')
|
||||
# self.start_state()
|
||||
self.convergence_start()
|
||||
return target
|
||||
|
||||
## heat PID
|
||||
|
Loading…