SLS Detector Package Patch Release 6.1.2 released on 25.11.2022
===============================================================
This document describes the differences between v6.1.2 and v6.1.1.
CONTENTS
--------
1. Resolved Issues
2. On-board Detector Server Compatibility
2. Firmware Requirements
3. Kernel Requirements
4. Download, Documentation & Support
1. Resolved Issues
==================
Detector Server
---------------
1. [Jungfrau, Moench, Ctb]
Programming Firmware
Updating kernel was not enough to ensure seamless programming of FPGA.
nCe signal now made high before programming blackfin.
Updating the on-board detector server to this version is recommended
before updating firmware for future releases.
Also removed the unnecessary test to check if it is allowed to update
kernel or on-board detector server, depending on the current kernel
version.
2. [Eiger] locking for start and stop
Since the start and stop has a few commands within the server itself,
they are locked to ensure they dont disturb the other to have undefined
behavior.
3. [Eiger] Quad Trimbits or threshold
Loading trimbits or setting threshold will throw an incorrect exception
for a quad as the left and right register values will not match. The fix
only sets and monitors right fpga for chip trim signals in quad.
4. [Eiger] Minor refactoring
Inconsistent reads between left and right register for read/write
register commands now give clear error messages.
Fixed some minor functions returned empty error messages when failing.
Client
------
1. [Eiger] Stop command freezes server
It is apparent in large detectors.
For example, if module A got a "stop" during an acquisition,
it stops to an 'Idle' state. If module B gets a "stop" before an
acquisition, it will return it is in 'Idle' state and continue
to start the next acquisition, but module B then waits for "ready
for trigger" synchronization from module A, which it will never get.
Since module B missed the asynchronous stop command, the workaround
in the client is to send another "stop" command (upto 10 retries)
till it returns Idle.
2. [Eiger][Jungfrau][Moench][Ctb]
Stop results in different frame numbers
A "stop" command will also now check for inconsistent 'nextframenumber'
between the modules and set it to the max + 1 to ensure that they
all start with the same frame number for the next acquisition.
2. On-board Detector Server Compatibility
==========================================
Eiger 6.1.2
Jungfrau 6.1.2
Gotthard 6.1.0
Mythen3 6.1.0
Gotthard2 6.1.0
Moench 6.1.2
Ctb 6.1.2
On-board Detector Server Upgrade
================================
From v6.1.0 (without tftp):
Using command 'updatedetectorserver'
From 5.0.0 (with tftp):
Using command 'copydetectorserver'
Instructions available at
https://slsdetectorgroup.github.io/devdoc/serverupgrade.html
3. Firmware Requirements
========================
Note: No change to 6.1.1.
Eiger
=====
Compatible version : 08.10.2021 (v29)
Jungfrau
========
Compatible version : 31.08.2021 (v1.2, PCB v1.0)
: 08.10.2021 (v2.2, PCB v2.0)
Gotthard
========
Compatible version : 08.02.2018 (50um and 25um Master)
: 09.02.2018 (25 um Slave)
Mythen3
=======
Compatible version : 10.09.2021 (v1.1)
Gotthard2
=========
Compatible version : 27.05.2021 (v0.1)
Moench
======
Compatible version : 05.10.2020 (v1.0)
Ctb
===
Compatible version : 05.10.2020 (v1.0)
Detector Firmware Upgrade
=========================
The following can be upgraded remotely:
Eiger via bit files
Jungfrau via 'programfpga' command <.pof>
Mythen3 via 'programfpga' command <.rbf>
Gotthard2 via 'programfpga' command <.rbf>
Moench via 'programfpga' command <.pof>
Ctb via 'programfpga' command <.pof>
The following cannot be upgraded remotely:
Gotthard
Instructions available at
https://slsdetectorgroup.github.io/devdoc/firmware.html
4. Kernel Requirements
======================
Note: No change to 6.1.1.
Blackfin
========
Latest version: Fri Oct 29 00:00:00 2021
Older ones will work, but might have issues with programming firmware via
the package.
Nios
====
Compatible version: Mon May 10 18:00:21 CEST 2021
Kernel Upgrade
==============
Eiger via bit files
Others via command
Commands: udpatekernel, kernelversion
Instructions available at
https://slsdetectorgroup.github.io/devdoc/commandline.html
https://slsdetectorgroup.github.io/devdoc/detector.html
https://slsdetectorgroup.github.io/devdoc/pydetector.html
5. Download, Documentation & Support
====================================
Download
--------
The Source Code:
https://github.com/slsdetectorgroup/slsDetectorPackage
Documentation
-------------
Installation:
https://slsdetectorgroup.github.io/devdoc/installation.html#
Firmware Upgrade:
https://slsdetectorgroup.github.io/devdoc/firmware.html
Detector Server upgrade:
https://slsdetectorgroup.github.io/devdoc/serverupgrade.html
Detector Simulators:
https://slsdetectorgroup.github.io/devdoc/virtualserver.html
Consuming slsDetectorPackage:
https://slsdetectorgroup.github.io/devdoc/consuming.html
API Examples:
https://github.com/slsdetectorgroup/api-examples
Command Line Documentation:
https://slsdetectorgroup.github.io/devdoc/commandline.html
Quick Start Guide:
https://slsdetectorgroup.github.io/devdoc/quick_start_guide.html
C++ API Documentation:
https://slsdetectorgroup.github.io/devdoc/detector.html
C++ API Example:
https://slsdetectorgroup.github.io/devdoc/examples.html#
Python API Documentation:
https://slsdetectorgroup.github.io/devdoc/pygettingstarted.html
Python API Example:
https://slsdetectorgroup.github.io/devdoc/pyexamples.html
Receivers (including custom receiver):
https://slsdetectorgroup.github.io/devdoc/receivers.html
https://slsdetectorgroup.github.io/devdoc/slsreceiver.html
TroubleShooting:
https://slsdetectorgroup.github.io/devdoc/troubleshooting.html
Further Documentation:
https://www.psi.ch/en/detectors/documentation
Info on Releases:
https://www.psi.ch/en/detectors/software
Support
-------
dhanya.thattil@psi.ch
erik.frojdh@psi.ch