802: jungfrau HW 1.0: adc output clock phase to 120 (#951)

* jungfrau: change adc output clock phase from 180 to 120 for v1.0 boards for reliable readout of adc #2

* formatting
This commit is contained in:
2024-08-22 14:27:16 +02:00
committed by GitHub
parent 7a28ba03c9
commit 8ada7b6e37
8 changed files with 25 additions and 2 deletions

View File

@ -1 +0,0 @@
../slsDetectorServers/jungfrauDetectorServer/bin/jungfrauDetectorServerv8.0.0

View File

@ -0,0 +1 @@
../slsDetectorServers/jungfrauDetectorServer/bin/jungfrauDetectorServerv8.0.2