mirror of
https://github.com/slsdetectorgroup/slsDetectorPackage.git
synced 2025-04-23 23:10:02 +02:00
rate capability plots
This commit is contained in:
parent
96d7345124
commit
6a0d44e358
@ -656,6 +656,19 @@ sls_detector_put 0-ratecorr -1
|
|||||||
|
|
||||||
Every time either the rate corrections are activated, $\tau$ is changed or the subframe length is changed, then a new correction table is evaluated. Note that computing the correction table is time consuming.
|
Every time either the rate corrections are activated, $\tau$ is changed or the subframe length is changed, then a new correction table is evaluated. Note that computing the correction table is time consuming.
|
||||||
|
|
||||||
|
Here in figure~\ref{rateplots} you can find typical values of $\tau$ and typical values of the rates for which we have small non linerities (10\% non linearity) as a function of the beam energy.
|
||||||
|
\begin{figure}[t]
|
||||||
|
\begin{center}
|
||||||
|
\includegraphics[width=.7\textwidth]{tauvsE}
|
||||||
|
\includegraphics[width=.7\textwidth]{Ratecapabilityflux}
|
||||||
|
\end{center}
|
||||||
|
\caption{Typical values of the dead time for the detector and safe fluxes as a function of the beam energy}.
|
||||||
|
\label{rateplots}
|
||||||
|
\end{figure}
|
||||||
|
|
||||||
|
|
||||||
|
|
||||||
|
|
||||||
\section{Dependent parameters and limits}
|
\section{Dependent parameters and limits}
|
||||||
Here is a list of dependent parameters:
|
Here is a list of dependent parameters:
|
||||||
\begin{enumerate}
|
\begin{enumerate}
|
||||||
|
BIN
manual/manual-client/Ratecapabilityflux.png
Normal file
BIN
manual/manual-client/Ratecapabilityflux.png
Normal file
Binary file not shown.
After Width: | Height: | Size: 16 KiB |
BIN
manual/manual-client/tauvsE.png
Normal file
BIN
manual/manual-client/tauvsE.png
Normal file
Binary file not shown.
After Width: | Height: | Size: 11 KiB |
Loading…
x
Reference in New Issue
Block a user