ctb: power regulators (set dac, get dac, get adc of voltage and current) works
This commit is contained in:
@ -1,6 +1,7 @@
|
||||
#pragma once
|
||||
|
||||
#include "commonServerFunctions.h" // blackfin.h, ansi.h
|
||||
#include "common.h"
|
||||
|
||||
/* AD7689 ADC DEFINES */
|
||||
|
||||
@ -62,8 +63,10 @@
|
||||
#define AD7689_ADC_DATA_NUMBITS (16)
|
||||
#define AD7689_NUM_CHANNELS (7)
|
||||
#define AD7689_NUM_INVALID_CONVERSIONS (3)
|
||||
|
||||
#define AD7689_INT_REF_MAX_MV (2500) // chosen using reference buffer selection in config reg
|
||||
#define AD7689_INT_REF_MIN_MV (0)
|
||||
#define AD7689_INT_MAX_STEPS (0xFFFF + 1)
|
||||
#define AD7689_TMP_C_FOR_1_MV (25.00 / 283)
|
||||
|
||||
|
||||
@ -144,17 +147,17 @@ int AD7689_GetTemperature() {
|
||||
AD7689_CFG_CFG_OVRWRTE_VAL);
|
||||
|
||||
// FIXME: do we have to read it 8 times?? (sequencer is disabled anyway) or are we sequencing, then we read only last channel
|
||||
uint16_t regval = AD7689_Get();
|
||||
int regval = AD7689_Get();
|
||||
|
||||
// value in mV FIXME: page 17? reference voltage temperature coefficient or t do with -40 to 85 °C
|
||||
uint16_t vmin = AD7689_INT_REF_MIN_MV;
|
||||
uint16_t vmax = AD7689_INT_REF_MAX_MV;
|
||||
uint16_t nsteps = (2 ^ AD7689_ADC_DATA_NUMBITS);
|
||||
uint16_t retval = (vmin + (vmax - vmin) * regval / (nsteps - 1));
|
||||
int retval = 0;
|
||||
ConvertToDifferentRange(0, AD7689_INT_MAX_STEPS,
|
||||
AD7689_INT_REF_MIN_MV, AD7689_INT_REF_MAX_MV,
|
||||
regval, &retval);
|
||||
FILE_LOG(logDEBUG1, ("\tvoltage read for temp: 0x%d mV\n", retval));
|
||||
|
||||
// value in °C
|
||||
uint16_t temp = AD7689_TMP_C_FOR_1_MV * retval;
|
||||
int temp = AD7689_TMP_C_FOR_1_MV * retval;
|
||||
FILE_LOG(logDEBUG1, ("\ttemp read: 0x%d °C\n", temp));
|
||||
|
||||
return temp;
|
||||
@ -191,13 +194,13 @@ int AD7689_GetChannel(int ichan) {
|
||||
AD7689_CFG_CFG_OVRWRTE_VAL);
|
||||
|
||||
// FIXME: do we have to read it 8 times?? (sequencer is disabled anyway) or are we sequencing, then we read only last channel
|
||||
uint16_t regval = AD7689_Get();
|
||||
int regval = AD7689_Get();
|
||||
|
||||
// value in mV
|
||||
uint16_t vmin = AD7689_INT_REF_MIN_MV;
|
||||
uint16_t vmax = AD7689_INT_REF_MAX_MV;
|
||||
uint16_t nsteps = (2 ^ AD7689_ADC_DATA_NUMBITS);
|
||||
uint16_t retval = (vmin + (vmax - vmin) * regval / (nsteps - 1));
|
||||
int retval = 0;
|
||||
ConvertToDifferentRange(0, AD7689_INT_MAX_STEPS,
|
||||
AD7689_INT_REF_MIN_MV, AD7689_INT_REF_MAX_MV,
|
||||
regval, &retval);
|
||||
FILE_LOG(logINFO, ("\tvoltage read for chan %d: 0x%d mV\n", retval));
|
||||
|
||||
return retval;
|
||||
@ -210,7 +213,7 @@ void AD7689_Configure(){
|
||||
FILE_LOG(logINFOBLUE, ("Configuring AD7689 (Slow ADCs): \n"));
|
||||
|
||||
// from power up, 3 invalid conversions
|
||||
FILE_LOG(logINFO, ("3 times due to invalid conversions from power up\n"));
|
||||
FILE_LOG(logINFO, ("\tConfiguring %d x due to invalid conversions from power up\n", AD7689_NUM_INVALID_CONVERSIONS));
|
||||
int i = 0;
|
||||
for (i = 0; i < AD7689_NUM_INVALID_CONVERSIONS; ++i) {
|
||||
AD7689_Set(
|
||||
|
Reference in New Issue