diff --git a/manual/manual-client/Eiger_short.tex b/manual/manual-client/Eiger_short.tex index 04fef6f8b..8e35824aa 100644 --- a/manual/manual-client/Eiger_short.tex +++ b/manual/manual-client/Eiger_short.tex @@ -22,7 +22,7 @@ \subsection{Mandatory setup - Hardware} An EIGER single module (500~kpixels) needs: \begin{itemize} -\item A chilled (water+alcohol) at approximately 21~$^{\circ}$C, which needs to dissipate 85~W. +\item A chilled (water+alcohol) at approximately 21~$^{\circ}$C, which needs to dissipate 85~W. For the (M, a special cooling liquid is required: 2/3 deionized water and 1/3 ESA Type 48. \item A power supply (12~V, 8~A). For the 9~M, a special cpu is give to remotely switch on and off the detector: see section~\ref{bchip100}. \item 2$\times$1~Gb/s Ethernet connectors to control the detector and, optionally, receive data at low rate. A DHCP server that gives IPs to the 1~Gb/s connectors of the detector is needed. Note that flow control has to be enabled on the switch you are using. \item 2$\times$10~Gb/s transceivers to optionally, receive data at high rate. @@ -57,6 +57,19 @@ The directory contains some executables taht are needed to make your detector to ./temp returns the water temperature returned by the flowmeter \end{verbatim} A watchdog is running on bchip100 to check for the flow and temparature. If outside of parameters, the detector will be switched off. +Here is an explanation of the LED color scheme of teh bchip100: +\begin{itemize} +\item NO LED Main Power off or Blackfin not ready, yet. +\item RED Too high temperature or too less water flow + Detector is shutted down and locked. + Detector will be unlocked (YELLOW) automatically when conditions are good again. +\item YELLOW Detector is off and unlocked. Ready to be turned on. +\item GREEN Detector is on +\end{itemize} +You can also Check temperatures and water flow in a browser (from teh same subnet where the 9M is: http://bchip100/status.cgi + + + \subsection{Mandatory setup - Receiver}