61 lines
2.8 KiB
Plaintext
61 lines
2.8 KiB
Plaintext
scripts and setup to run version 1 of the jungfrau calibration
|
|
CuFDBCS = copper fluorescence, direct beam, current source
|
|
|
|
Module 006 - ESRF
|
|
./JFMC_CuFluoPeak 006 /mnt/pcmoench_jungfrau_data/Module_006_161116 10us_500Hz_QS_allgain_pede 10us_500Hz_QS_G0_Cu
|
|
./JFMC_CuFluoPeakFit 006
|
|
./JFMC_DirectBeamScan 006 /mnt/pcmoench_jungfrau_data/Module_006_161116 100us_500Hz_QS_allgain_pede_1055 100us_500Hz_QS_beam_scan_a
|
|
./JFMC_CurrentSourceScan 006 /mnt/pcmoench_jungfrau_data/Module_006_161116 Current_source_scan_a
|
|
./JFMC_CalibWriter 006
|
|
|
|
|
|
Module 008 - Pohang
|
|
./JFMC_CuFluoPeak 008 /data_pool/Module_008_181116 10us_500Hz_QS_allgain_pede_1558 10us_500Hz_QS_G0_Cu_2
|
|
./JFMC_CuFluoPeakFit 008
|
|
./JFMC_DirectBeamScan 008 /mnt/pcmoench_jungfrau_data/Module_008_181116 150us_500Hz_QS_allgain_pede_1332 150us_500Hz_QS_beam_scan_b
|
|
./JFMC_CurrentSourceScan 008 /mnt/pcmoench_jungfrau_data/Module_008_181116 Current_source_scan_a
|
|
./JFMC_CalibWriter 008
|
|
|
|
|
|
Module 021 - 1M top LCLS, PX
|
|
./JFMC_CuFluoPeak 021 /data_pool/1M_021_022_150916 10us_200Hz_QS_allgain_pede__B 10us_200Hz_QS_G0_Cu__B
|
|
./JFMC_CuFluoPeakFit 021
|
|
./JFMC_DirectBeamScan 021 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 200us_200Hz_QS_allgain_pede_scan_d_B 200us_200Hz_QS_beam_scan_d_B
|
|
./JFMC_CurrentSourceScan 021 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 Current_source_scan_a_B
|
|
./JFMC_CalibWriter 021
|
|
|
|
|
|
Module 022 - 1M bottom LCLS, PX
|
|
./JFMC_CuFluoPeak 022 /data_pool/1M_021_022_150916 10us_200Hz_QS_allgain_pede__A 10us_200Hz_QS_G0_Cu__A
|
|
./JFMC_CuFluoPeakFit 022
|
|
./JFMC_DirectBeamScan 022 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 200us_200Hz_QS_allgain_pede_scan_d_A 200us_200Hz_QS_beam_scan_d_A
|
|
./JFMC_CurrentSourceScan 022 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 Current_source_scan_a_A
|
|
./JFMC_CalibWriter 022
|
|
|
|
|
|
Module 027 - old HDI (no current source scan). sent to ESRF with Paolo
|
|
./JFMC_CuFluoPeak 027 /data_pool/Module_027_Calib FCu_40kvp_60mA_10us_500Hz_QS_pede FCu_40kvp_60mA_10us_500Hz_QS_beam
|
|
|
|
|
|
Module 040 - partly etched. sent to ESRF with Paolo
|
|
./JFMC_CuFluoPeak 040 /data_pool/Module_040_Calib Cu_FLuo_240317_500Hz_10us_QS_pede Cu_FLuo_240317_500Hz_10us_QS_beam
|
|
./JFMC_CurrentSourceScan 040 /data_pool/Module_040_Calib CS_240322_scan
|
|
|
|
|
|
Module 044 - added to the 1M to make the 1.5M, cSAXS
|
|
./JFMC_CuFluoPeak 044 /data_pool/Module044 allpede_10us_500Hz_QS_G0_1500 CuFluo_10us_500Hz_QS_G0
|
|
./JFMC_CuFluoPeakFit 044
|
|
./JFMC_DirectBeamScan 044 /data_pool/Module044 allpede_150us_500Hz_QS_1620 DB_170221_scan1
|
|
./JFMC_CurrentSourceScan 044 /data_pool/Module044 CS_170221_scan2
|
|
./JFMC_CalibWriter 044
|
|
|
|
|
|
Module 028
|
|
direct beam scan data only
|
|
./JFMC_DirectBeamScan 028 /mnt/pcmoench_jungfrau_data/test_module028_direct_beam_scan__01092016 200us_scan_a_pedestal_QS 200us_scan_a_beam_QS
|
|
|
|
|
|
Module 032
|
|
current source scan data only
|
|
./JFMC_CurrentSourceScan 032 /mnt/pcmoench_jungfrau_data/test_module032_cal_source_15082016 test_currentsourcescan_5
|