be included in the average. Reran everything, looks same or better. Also removed zoom plot, unnecessary. Repointed data files to new storage name.
70 lines
3.4 KiB
Plaintext
70 lines
3.4 KiB
Plaintext
scripts and setup to run version 1 of the jungfrau calibration
|
|
CuFDBCS = copper fluorescence, direct beam, current source
|
|
|
|
Module 006 - ESRF
|
|
./JFMC_CuFluoPeak 006 /mnt/pcmoench_jungfrau_data/Module_006_161116 10us_500Hz_QS_allgain_pede 10us_500Hz_QS_G0_Cu
|
|
./JFMC_CuFluoPeakFit 006
|
|
./JFMC_DirectBeamScan 006 /mnt/pcmoench_jungfrau_data/Module_006_161116 100us_500Hz_QS_allgain_pede_1055 100us_500Hz_QS_beam_scan_a
|
|
./JFMC_CurrentSourceScan 006 /mnt/pcmoench_jungfrau_data1_ib/Module_006_161116 Current_source_scan_a
|
|
./JFMC_CalibWriter 006
|
|
|
|
|
|
Module 008 - Pohang
|
|
./JFMC_CuFluoPeak 008 /mnt/pcmoench_jungfrau_data/Module_008_181116 10us_500Hz_QS_allgain_pede_1558 10us_500Hz_QS_G0_Cu_2
|
|
./JFMC_CuFluoPeakFit 008
|
|
./JFMC_DirectBeamScan 008 /mnt/pcmoench_jungfrau_data/Module_008_181116 150us_500Hz_QS_allgain_pede_1332 150us_500Hz_QS_beam_scan_b
|
|
./JFMC_CurrentSourceScan 008 /mnt/pcmoench_jungfrau_data1_ib/Module_008_181116 Current_source_scan_a
|
|
./JFMC_CalibWriter 008
|
|
|
|
|
|
Module 021 - 1M top LCLS, PX
|
|
./JFMC_CuFluoPeak 021 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 10us_200Hz_QS_allgain_pede__B 10us_200Hz_QS_G0_Cu__B
|
|
./JFMC_CuFluoPeakFit 021
|
|
./JFMC_DirectBeamScan 021 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 200us_200Hz_QS_allgain_pede_scan_d_B 200us_200Hz_QS_beam_scan_d_B
|
|
./JFMC_CurrentSourceScan 021 /mnt/pcmoench_jungfrau_data1_ib/1M_021_022_150916 Current_source_scan_a_B
|
|
./JFMC_CalibWriter 021
|
|
|
|
|
|
Module 022 - 1M bottom LCLS, PX
|
|
./JFMC_CuFluoPeak 022 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 10us_200Hz_QS_allgain_pede__A 10us_200Hz_QS_G0_Cu__A
|
|
./JFMC_CuFluoPeakFit 022
|
|
./JFMC_DirectBeamScan 022 /mnt/pcmoench_jungfrau_data/1M_021_022_150916 200us_200Hz_QS_allgain_pede_scan_d_A 200us_200Hz_QS_beam_scan_d_A
|
|
./JFMC_CurrentSourceScan 022 /mnt/pcmoench_jungfrau_data1_ib/Module_022_Calib CS_HS_NF_200Hz_scan1
|
|
./JFMC_CalibWriter 022
|
|
|
|
|
|
Module 027 - old HDI (no current source scan). sent to ESRF with Paolo
|
|
./JFMC_CuFluoPeak 027 /mnt/pcmoench_jungfrau_data/Module_027_Calib FCu_40kvp_60mA_10us_500Hz_QS_pede FCu_40kvp_60mA_10us_500Hz_QS_beam
|
|
./JFMC_CuFluoPeakFit 027
|
|
./dummyRootFileMaker
|
|
|
|
|
|
Module 040 - partly etched. sent to ESRF with Paolo
|
|
./JFMC_CuFluoPeak 040 /mnt/pcmoench_jungfrau_data/Module_040_Calib Cu_FLuo_240317_500Hz_10us_QS_pede Cu_FLuo_240317_500Hz_10us_QS_beam
|
|
./JFMC_CuFluoPeakFit 040
|
|
./JFMC_CurrentSourceScan 040 /mnt/pcmoench_jungfrau_data1_ib/Module_040_Calib CS_240322_scan
|
|
|
|
|
|
Module 044 - added to the 1M to make the 1.5M, cSAXS
|
|
./JFMC_CuFluoPeak 044 /mnt/pcmoench_jungfrau_data/Module044 allpede_10us_500Hz_QS_G0_1500 CuFluo_10us_500Hz_QS_G0
|
|
./JFMC_CuFluoPeakFit 044
|
|
./JFMC_DirectBeamScan 044 /mnt/pcmoench_jungfrau_data/Module044 allpede_150us_500Hz_QS_1620 DB_170221_scan1
|
|
./JFMC_CurrentSourceScan 044 /mnt/pcmoench_jungfrau_data1_ib/Module044 CS_170221_scan2
|
|
./JFMC_CalibWriter 044
|
|
|
|
|
|
Module 045 - replaces M021 in the 1M
|
|
./JFMC_CuFluoPeak 045 /mnt/pcmoench_jungfrau_data/Module_045_Calib pede_G0_1649_HS_NF_500Hz Cu_HS_NF_500Hz
|
|
./JFMC_CuFluoPeakFit 045
|
|
./JFMC_CurrentSourceScan 045 /mnt/pcmoench_jungfrau_data1_ib/Module_045_Calib CS_HS_NF_200Hz_scan10
|
|
|
|
|
|
Module 028
|
|
direct beam scan data only
|
|
./JFMC_DirectBeamScan 028 /mnt/pcmoench_jungfrau_data/test_module028_direct_beam_scan__01092016 200us_scan_a_pedestal_QS 200us_scan_a_beam_QS
|
|
|
|
|
|
Module 032
|
|
current source scan data only
|
|
./JFMC_CurrentSourceScan 032 /mnt/pcmoench_jungfrau_data1_ib/test_module032_cal_source_15082016 test_currentsourcescan_5
|