feat(LamNI): add automated rotation-center calibration via X-ray eye
lsamx_center/lsamy_center have to be re-measured by hand for nearly every
new sample (the rotation stage is tilted on top of lsamx/lsamy, so the
effective axis position shifts with sample thickness/mounting). Add
XrayEyeAlign.find_rotation_center()/lamni.rotation_center_calibration_start()
to automate this using the existing X-ray eye GUI and lamni_move_to_scan_center
interferometer-feedback move:

- "isolated" sample: click particle centre at 0 and 180 deg, use the
  midpoint (angle-tilt-independent) as the rotation axis position.
- "extended" sample: live 0->180->0 sweep for visual identification, single
  click, then a verification sweep and an accept/iterate prompt.

Both write the result to lsamx/lsamy's "center" userParameter after
confirmation. Heavily verbose logging throughout since this is fundamentally
an interactive, hardware-in-the-loop procedure that can't be fully exercised
by automated tests.

Co-Authored-By: Claude Sonnet 5 <noreply@anthropic.com>