@@ -27,12 +27,16 @@ The recommended way to (re-)measure `center` for a new sample is the GUI-driven
|
||||
This opens the X-ray eye widget and walks through the calibration interactively. Pick the function based on what's mounted:
|
||||
|
||||
- `lamni.xrayeye_rotation_center_calibration_isolated()` — for a sparse/isolated particle. You submit its centre position once at 0° and once at 180°; the midpoint of the two is the rotation axis' projected position (this works regardless of the axis tilt). No further confirmation step.
|
||||
- `lamni.xrayeye_rotation_center_calibration_extended()` — for a non-isolated/textured sample where the rotation centre can be identified visually. The sample sweeps 0° → 180° → 0° with the live view left open so you can watch for the point that doesn't move, then you submit a single click at 0°. The correction is applied, a short verification sweep (0° → 45° → 0°) runs, and you're asked *"Are you happy with this alignment, or shall we do another iteration?"* — answer `n` to refine further.
|
||||
- `lamni.xrayeye_rotation_center_calibration_extended()` — for a non-isolated/textured sample where the rotation centre can be identified visually. The sample sweeps 0° → 180° → 0° with the live view left open so you can watch for the point that doesn't move, then you submit a single click at 0°. The correction is applied, a short verification sweep (0° → 45° → 0°) runs, and you're asked *"Alignment acceptable — stop here?"* — answer `n` to refine further.
|
||||
|
||||
At the end you're shown the computed `lsamx_center`/`lsamy_center` and asked to confirm before they're written via `dev.lsamx.update_user_parameter(...)` / `dev.lsamy.update_user_parameter(...)`.
|
||||
|
||||
Pass `keep_shutter_open=True` if it's hard to relocate the sample between steps, matching `xrayeye_alignment_start()` below. Interrupt with Ctrl-C to abort; the crosshair and DAP forwarding are cleaned up automatically.
|
||||
|
||||
**Experimental: continuous-rotation "smear" aid**
|
||||
|
||||
`lamni.xrayeye_rotation_center_calibration_smear_experimental(sweep_deg=360.0, keep_shutter_open=False)` is an **experimental** alternative to `..._extended()` for the same non-isolated-sample case. Instead of judging the rotation centre from a single instant of live rotation, it rotates `lsamrot` continuously through `sweep_deg` (default a full circle) while accumulating a max-projection ("star-trail") composite from the camera: off-axis features smear into circular arcs, and the common center of curvature of those arcs is the rotation axis — usually an easier target to click than one live frame. `sweep_deg` can be reduced below 360 (even below 180) if a shorter arc already shows enough curvature. There is no automatic circle fitting — you still submit the centre by eye, same click mechanism as `..._extended()`, and the same iterate/verify/apply flow follows. Not yet merged into the production branch; try it from `experimental/rotation_center_smear`.
|
||||
|
||||
**Manual fallback**
|
||||
|
||||
To observe the axis of rotation obtain the position of the Fresnel zone plate on the X-ray eye, possibly in the *ueye gui* by:
|
||||
|
||||